Item General Specifications
Dimensions and Mass
Withoutprotector(option010)
Size:270(W)×165(H)×61(D)mm
10.6×6.5×2.4inches
Weight:2.2kg(4.8lbs)includingbattery
Withprotector(option010)
Size:284(W)×200(H)×77(D)mm
11.2×7.9×3inches
Weight:2.9kg(6.4lbs)includingbattery
Display
6.5inchTFT-LCD(640×480,withbacklight,transparenttype),enhancedindoor/outdooroptional
Interface USB1.1,TypeA×1(memory),TypeB×1(USBmassstorage)
Data Storage
Internalmemory:440MB(upto1000traces),
Externalmemory(USB):upto30,000traceswith512MB
PowerSupply
12VDC,
100to240VAC,Allowableinputvoltagerange:90to264V,50/60Hz
Battery
Type:Lithiumion
OperatingTime
∗
1
:8hours
RechargeTime:<5hours(poweroff)
PowerSavingFunctions
Backlightoff:Disable/1to99minutes
Autoshutdown:Disable/1to99minutes
VerticalScale 0.13,0.33,0.65,1.3,3.25,6.5,13dB/div
IORSetting 1.400000to1.699999(0.000001steps)
Units km,m,kft,ft,mi
Languages
Userselectable(English,SimplifiedChinese,TraditionalChinese,French,German,Italian,Korean,Portuguese,
Russian,SpanishandSwedish-contactAnritsuforavailabilityofothers)
SamplingPoints
∗
2
Normal:5001,Highdensity:20001or25001(MT9083AandBseries),upto150,001(MT9083Cseries)
SamplingResolution
∗
3
5cm(min.)
ReflectanceAccuracy Singlemode:±2dB,multimode:±4dB
DistanceAccuracy ±1m±3×measurementdistance×10
-5
±markerresolution(excludingIORuncertainty)
Distance Range
Singlemode:0.5,1,2.5,5,10,25,50,100,200km(MT9083AandBseries,except780nm:0.5,1,2.5km)
0.5,1,2.5,5,10,25,50,100,200,300km(MT9083Cseries)
Multimode:0.5,1,2.5,5,10,25,50,100km
Testing Modes
Faultlocate:Providesend/breaklocation,endtoendloss,fiberlength
StandardOTDR:Userselectableautomaticormanualset-up
ConstructionOTDR:Automated,multi-wavelengthtesting
Lightsource:StabilizedLightsource(CW,270Hz,1kHz,2kHzoutput)
Losstestset(optional):PowermeterandLightsource
ConnectorVideoInspectionProbe
Visualfaultlocator(optional):Visibleredlightforfiberidentificationandtroubleshooting
FiberEventAnalysis
Autoormanualoperation,displayedintableformat
UserdefinedPASS/FAILthresholds:
-Reflectiveandnon-reflectiveevents:0.01to9.99dB(0.01dBsteps)
-Reflectance:–70.0to–20.0dB(0.1dBsteps)
-Fiberend/break:1to99dB(1dBsteps)
Numberofdetectedevents:upto99
Macrobend detection
OTDR Trace Format Telcordiauniversal.SOR,issue2(SR-4731)
Other Functions
Realtimesweep
∗
4
:0.15sec.
Lossmodes:2pointloss,dB/km,2pointLSA,spliceloss,ORL
Averagingmodes:Timed(1to3600sec.)
LiveFiberdetect:Verifiespresenceofcommunicationlightinopticalfiber
Connectioncheck:AutomaticcheckofOTDRtoFUTconnectionquality
Traceoverlayandcomparison,Templatefunction,USBkeyboardsupport,Remotecontrol,VideooutputtoPC
EnvironmentalConditions
Operatingtemperatureandhumidity:0°to+40°C(MT9083AandBseries),<80%(non-condensing)
0°to+45°C(MT9083Cseries),<80%(non-condensing)
Storagetemperatureandhumidity:–20°to+60°C,<80%(non-condensing)
Vibration:ConformingtoMIL-T-28800EClass3
Dustproof:MIL-T-28800EClass2
Dripproof:IP51(IEC60529),JISC0920TYPEI
EMC EN61326-1,EN61000-3-2
LVD EN61010-1
Notes
∗1:Typical,backlightoff,sweepinghaltedat25˚C,6hourstypicalcontinuoustesting
∗2:Eitherhighdensityvalueisselecteddependingondistancerange
∗3:Except780nm
∗4:Resolution:LowDensity.Exceptmodels062,068-1sec.orless
Specifications
MT9083A/B/C ACCESS Master Common Specifications
9
Commenti su questo manuale